When the WS425 A/B NMEA Standard profile is configured, WMT700 sends messages based on the configured automatic message interval. There are no operating commands available for the user.
The following table lists configurable parameters and their allowed and default values for the WS425 A/B NMEA Standard profile
Parameter | Default value | Profile-specific allowed values | Description |
---|---|---|---|
address | 1 | String with 1 character | Address for WMT700 |
autoInt | 1 |
0.125 ... 1000 Resolution: 0.125 |
Automatic message interval in seconds. Do not select a message interval that is shorter than the time it takes to send a data message |
autoPort | 1 |
1 = COM1 port 2 = COM2 port |
Serial port to which WMT700 sends automatic data messages |
autoSend | 0 |
0 = Automatic messages disabled 19 = NMEA automatic data message |
Automatic data message number. Selects the data message format for automatic messages |
com1_protocol | 0 1 | 4 = WS425 A/B NMEA Standard | Profile for serial port COM1 |
com2_protocol | 0 | 4 = WS425 A/B NMEA Standard | Profile for serial port COM2 |
com2_interf | N/A 2 |
0 = RS-485 1 = RS-422 3 = RS-232 |
Interface for serial port COM2 |
wndAvg | 1 |
0.125 … 3600 Resolution: 0.125 |
Averaging time for wind measurement in seconds |
wndOrientation | 0 |
0 = Array facing up 1 = Array facing down |
Orientation of the array of WMT700 |
wndUnit | 0 2 |
0 = Meters per second 1 = Miles per hour 2 = Kilometers per hour 3 = Knots |
Wind speed unit |
wndVector | 1 | 0 = Scalar averaging | Wind averaging method |
1 The parameter
has no protocol-specific default value. It is determined separately in the
configuration code.
2 Wind speed
unit and COM2 digital communication interface are defined in the order form. Default
setting may be any of the allowed options, and it can be verified from the unit’s
configuration code.