WD - report status of built-in tests - WMT700

WMT700 Series User Guide

Document code
M211095EN
Revision
K
Language
English
Product
WMT700
Document type
User guide

The WD command reports the status of the built-in tests (BIT). The command offers a concise report on the pass/fail status of all BITs. The command format is:

WD CR LF
The response:
STX WD PPPPPPPPPPPP CS ETX CR LF
Name Description
STX ASCII start of test control character (position 1)
WD Command identifier (positions 2,3)
P or F Pass/Fail for 12 built in tests specified in WS command above in the order shown in the WS command text (positions 4-15) except Thermistor temperature.
CS Modulo 256 hexadecimal checksum of all characters from position 2 through the last P or F (positions 16,17)
EXT ASCII end of text control character (position 18)
CR ASCII carriage return control character (position 19)
LF ASCII line feed control character (position 20)