The WT command requests the short response built-in test (BIT) results and status. The command format:
WT CR LF
The response:
Name | Description |
---|---|
Byte 1 | STX start of text ASCII control character |
Byte 2 | W sensor ID |
Byte 3 | T command identifier |
Byte 4 | P or F for heater voltage pass or fail |
Byte 5-8 | XX.X format heater voltage (volts RMS) |
Byte 9 | P or F for array heater resistance pass or fail |
Byte 10-12 | X.X or XX. (format as required) for array heater resistance (ohms) |
Byte 13 | P or F for pass or fail on heater off voltage |
Byte 14-16 | X.X format heater off voltage (volts RMS) |
Byte 17 | P or F for pass or fail thermistor temperature |
Bytes 18-20 | Thermistor temperature (degrees Celsius) |
Byte 21 | P or F for incoming supply voltage pass or fail |
Byte 22-25 | XX.X format incoming supply voltage (volts DC) |
Byte 26 | P or F for 5 volt supply pass or fail |
Bytes 26-29 | X.XX format 5 volt supply voltage (volts DC) |
Byte 30 | P or F for 10 volt supply pass or fail |
Bytes 31-34 | XX.X format 10 volt supply voltage (volts DC) |
Byte 35 | ASCII hyphen character (-) |
Bytes 36-40 | XXXXX format bad 1 second reading counter |
Byte 41 | P or F for path 0 Signal Quality Index pass or fail |
Bytes 42-43 | XX format path 0 Signal Quality Index |
Byte 44 | P or F for path 1 Signal Quality Index pass or fail |
Bytes 45-46 | XX format path 1 Signal Quality Index |
Byte 47 | P or F for path 2 Signal Quality Index pass or fail |
Bytes 48-49 | XX format path 2 Signal Quality Index |
Byte 50 | P or F for path 3 Signal Quality Index pass or fail |
Bytes 51-52 | XX format path 3 Signal Quality Index |
Byte 53 | P or F for path 4 Signal Quality Index pass or fail |
Bytes 54-55 | XX format path 4 Signal Quality Index |
Byte 56 | P or F for path 5 Signal Quality Index pass or fail |
Bytes 57-58 | XX format path 5 Signal Quality Index |
Bytes 59-60 | XX hexadecimal ASCII for modulo 256 checksum for Character positions 2 through 58 |
Byte 61 | ETX end of text ASCII control character |
Byte 62 | ASCII carriage return control character |
Byte 63 | ASCII line feed control character |