aV! — start verification - WMT700

WMT700 Series User Guide

Document code
M211095EN
Revision
K
Language
English
Product
WMT700
Document type
User guide

The aV! command starts verification. However, WMT700 does not return verification data directly after this command. It sends a response that specifies when the verification results can be retrieved using the aD0! command and the number of parameters. The WMT700 verification data includes a watchdog count, sensor measurement unit settings, and the current SDI mode setting.

The command:

aV!
Name Description
a Single-digit WMT700 address that corresponds to the first character of the address value
V Starts verification command
! Terminates the command

The response:

atttn<CR><LF>
Name Description
a Single-digit WMT700 address that corresponds to the first character of the address value
ttt Period of time, in seconds, after which WMT700 has the verification data ready
n Number of returned verification data fields
<CR><LF> Terminates the response

Example

1V! 10014<CR><LF>

Interpretation of the example message:

  • Time when the measurement is ready: 1 second
  • Number of returned measurement values: 4

To fetch the measurement results, use the aD0! command.