When you have selected WS425 A/B NMEA Extended profile, you can set WMT700 to send messages based on the configured automatic message interval or poll data using the $WIP command.
Parameter | Default value | Profile-specific allowed values | Description |
---|---|---|---|
address | A | String with 1 character | Address for WMT700 |
autoInt | 1 |
0.125 ... 1000 Resolution: 0.125
|
Automatic message interval in seconds. Do not select a message interval that is shorter than the time it takes to send a data message |
autoPort | 2 |
1 = COM1 port 2 = COM2 port |
Serial port to which WMT700 sends automatic data messages |
autoSend | 15 |
0 = Automatic messages disabled 15 = NMEA automatic data message |
Automatic data message number. Selects the data message format for automatic messages. |
com1_protocol | 0 1 | 5 = WS425 A/B NMEA Extended | Profile for serial port COM1 |
com2_protocol | 5 | 5 = WS425 A/B NMEA Extended | Profile for serial port COM2 |
com2_interf | N/A 2 |
0 = RS-485 1 = RS-422 3 = RS-232 |
Interface for serial port COM2 |
wndAvg | 3 |
0.125 … 3600 Resolution: 0.125 |
Averaging time for wind measurement in seconds |
wndOrientation | 0 |
0 = Array facing up 1 = Array facing down |
Orientation of the array of WMT700 |
wndUnit | 0 2 |
0 = Meters per second 1 = Miles per hour 2 = Kilometers per hour 3 = Knots |
Wind speed unit |
wndVector | 0 | 0 = Scalar averaging | Wind averaging method |
1 COM1 service
port is always by default 0 - WMT700 protocol. This can be changed using serial
commands if COM1 is needed for protocol-specific communication
2 Wind speed
unit and COM2 digital communication interface are defined in the order form. Default
setting may be any of the allowed options, and it can be verified from the unit’s
configuration code.