Connect a cable between your terminal computer, power supply, and WMT700.
-
Open the HyperTerminal program.
-
Cancel the new connection.
-
Select .
-
Select the correct COM port and select Configure.
-
Set the communication parameters according to the configured port settings of WMT700.
The default values for WMT700 are:
Communication parameter |
Default value |
Bits per second |
9600 |
Data bits |
8 |
Parity |
None |
Stop bits |
1 |
Flow control |
None |
-
Select .
-
Select . Close the New Connection Properties window.
-
Select .
-
Select . Type a name and select an icon for the connection. Select
OK.
-
Switch the sensor power supply on. The following information is displayed:
WMT700 v<version number>
-
Wait for 4 seconds. WMT700 enters measurement mode automatically.
-
To test the connection, enter configuration mode with the following command:
$0OPEN<CR><LF>
-
Use the following command to return to measurement mode:
CLOSE<CR><LF>
You can now start operating WMT700. Note
that WMT700 does not send data messages automatically unless configured to do so.