Manage Devices

  1. Ensure no users are logged on to the HMT140 you want to add.
  2. In Sites Manager select the Hosts and Devices tab.
  3. On the Hosts and Devices tree select the Vaisala viewLinc Enterprise Server or a device host, then select Configure > Add Device.
  4. In the Add Device window, in the Device Class field, select HMT140.
  5. Complete the following:
    Timeout
    To ensure continuous monitoring, do not change (default 30 seconds).
    Serial number
    Type the HMT140 serial number.
    UDP port
    Can be modified if required.
    Max blocks per beacon
    Accept the default number of data blocks (64) transferred between viewLinc and the device to maximize network efficiency, or modify as required (256 max).
    Do not change the Max blocks per beacon value for HMT140 data loggers without first consulting your technical support department. Changes to this setting may adversely impact battery life.