Error codes in MI70 Handheld Meter - HMP110 - HMP113 - HMP60 - HMP63 - MI70

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide
Table 1. Error codes in MI70

If several errors are active at the same time, the MI70 shows the sum of currently active error codes. The label "hex" might not be displayed in MI70.

Refer to Error messages in Insight software for more information about the error states.

Transmitter error code in MI70 Corresponding error
1hex Temperature measurement error. [44]
2hex Humidity measurement error. [45]
4hex Humidity sensor failure. [46]
8hex Capacitance reference error. [47]
10hex Ambient temperature out of range. [48]
20hex Firmware checksum mismatch. [49]
40hex Device settings corrupted. [50]
80hex Additional configuration settings corrupted. [51]
100hex Sensor coefficients corrupted. [52]
200hex Main configuration settings corrupted. [53]
800hex Supply voltage out of range. [55]
2000hex Non-volatile memory read/write failure. [57]
4000hex Calibration certificate checksum mismatch. [58]