Use the FILT command to view or set the speed at which the latest measurement result is integrated into the humidity and temperature readings. The command affects both analog output and serial line output.

FILT [a.aaa]<cr>

Syntax Description
a.aaa

Range 0.001 ... 1.0.

1.0 = No filtering, the latest measurement is output without averaging

0.5 = Average of the last 2 measurements

0.1 = Average of approximately 16 measurements

Example (default setting, no filtering):

filt
Filter         :    1.000     ?

Example (set filtering to 0.5):

filt 0.5
Filter         :    0.500