Configuration registers - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide
Table 1. Modbus configuration data registers (writable)
Register number Register address Register description Data format Unit / Valid range
Filtering
785 0310hex Measurement filtering factor 32-bit float 0.001 (maximum filtering) … 1.000 (no filtering)
0311hex
Communication
1537 0600hex Modbus address 16-bit integer 1 … 247
1538 0601hex Bit rate 1 16-bit integer

5 = 9600

6 = 19200

7 = 38400

8 = 57600

1539 0602hex Parity, data, stop bits 1 16-bit integer

0 = None, 8, 1

1 = None, 8, 2

2 = Even, 8, 1

3 = Even, 8, 2

4 = Odd, 8, 1

5 = Odd, 8, 2

1540 0603hex Response delay 16-bit integer

0 … 1020 ms

1541 0604hex Protocol 16-bit integer

6 = Modbus RTU

1542 0605hex Restart device 16-bit integer

When writing to register: 1 = Restart the device

If incompatible settings/protocol are selected and the probe is then restarted, it might no longer be possible to communicate with the probe using Modbus. You can override invalid communication settings in Vaisala Insight software, Settings menu. See Vaisala Insight software.
1 Changed setting will be activated on device restart.