Overview

All HMP60 series probes and HMP110 series probes, with the exception of the HMP110REF reference probe, can be ordered with analog output channels. Probes that use analog output can also use the RS-485 interface for temporary service access (Vaisala Industrial Protocol serial line commands) and Modbus RTU communication.

For installations where permanent digital output is required, the HMP110 model can be ordered as a digital-only probe (select either Modbus, RS-485 or Interface for Vaisala devices (VDIGI) as the output when ordering).

All HMP60 series and HMP110 series probes can also be connected temporarily to Insight PC software with a Vaisala USB cable for configuration, calibration and adjustment, and monitoring.

When you order HMP110 as a digital probe, analog output channels are not available.

Analog output options

Table 1. HMP60 and HMP110 series analog output options
HMP60 HMP63

HMP110

Analog

HMP110

Digital

HMP110T HMP113 HMP110REF
Analog output channels 2 2 2 None 1 2 None

Voltage output

(0 ... 1/2.5/5 VDC,

1 ... 5 VDC)

Yes Yes Yes No Yes Yes No
Current output Yes 1 Yes 1 Yes 1 No Yes 1 Yes 1 No

Digital output options

For installations where permanent digital output is required, select the digital-only HMP110 model (digital output with Modbus, RS-485, or Interface for Vaisala devices (VDIGI), selected when ordering). The digital output options in other probe models are intended for temporary use.
Table 2. HMP60 and HMP110 series digital output options
HMP60 HMP63

HMP110

Analog

HMP110

Digital

HMP110T HMP113 HMP110REF
RS-485 digital output Temporary service access only Yes Temporary service access only
Modbus RTU Yes Yes Yes Yes Yes Yes Yes
Temporary Insight PC software configuration access with accessory USB cable Yes Yes Yes Yes Yes Yes Yes
1 4 … 20 mA with separate loop power converter, compatible with humidity accuracy only.