Property | Description/Value |
---|---|
Relative humidity | |
Measurement range | 0 … 100 %RH |
Typical accuracy: | |
at 0 … +40 °C (+32 … +140 °F) |
±3 %RH (0 … 90 %RH) ±5 %RH (90 … 100 %RH) |
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F and +104 … +140 °F) |
±5 %RH (0 … 90 %RH) ±7 %RH (90 … 100 %RH) |
Humidity sensor | Vaisala INTERCAP® |
Temperature | |
Measurement range | −40 … +60 °C (−40 … +140 °F) |
Accuracy: | |
at +10 … +30 °C (+50 … +86 °F) at −40 … +10 and +30 … +60 °C (−40 … +50 and +86 … +140 °F) |
±0.5 °C (±0.9 °F) ±0.6 °C (±1.08 °F) |
Dew point temperature | |
Measurement range | −40 … +60 °C (−40 … +140 °F) |
Typical accuracy: | |
at 0 … +40 °C (+32 … +140 °F) when dew point depression1 < 15 °C | ±2 °C (±3.6 °F) |
at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F and +104 … +140 °F) when dew point depression < 10 °C 1 |
±3 °C (±5.4 °F) |
Analog outputs | |
Accuracy at 20 °C (+68 °F) | ±0.2 % of FS |
Temperature dependence | ±0.01 % of FS/°C (±0.006 % of FS/°F) |
Measurement ranges | |
Dew point temperature | −40 … +60 °C (−40 … +140 °F) |
Wet bulb temperature | −40 … +60°C (−40 … +140°F) |
Absolute humidity | 0 … 130 g/m3 (0 … 57 gr/ft3) |
Mixing ratio | 0 … 153 g/kg (0 … 1069 gr/lb) |
Enthalpy | −40 … 459 kJ/kg (−9.6 … 207 BTU/lb) |
Property | Description/Value |
---|---|
Operating temperature | −40 … +60 °C (−40 … +140 °F) |
IP rating 2 |
IP54: Limited protection against dust. Protected from water splashes from any direction. |
Property | Description/Value |
---|---|
Power consumption | 1 mA average, max. peak 5 mA |
Operating voltage3 | |
With 1 V / 2.5 V output | 5 … 28 V DC |
With 5 V output | 8 … 28 V DC |
With loop-power converter | 8 … 28 V DC |
With digital output | 5 … 28 V DC |
Start-up time | |
Probes with analog output |
4 s at operating voltage 13.5 … 16.5 V DC 2 s at other valid operating voltages |
Probes with digital output | 1 s |
Outputs | |
2 channels | 0 … 1 V DC / 0 … 2.5 V DC / 0 … 5 V DC / 1 … 5 V DC |
1-channel loop-power converter (separate module, compatible with humidity accuracy only) | 4 … 20 mA |
Digital output (optional) | RS-485 2-wire half duplex, supports Modbus RTU |
External loads | |
0 … 1 V | RL min. 10 kΩ |
0 … 2.5 V /0 … 5 V | RL min. 50 kΩ |
Output parameters | |
Relative humidity, temperature, dew point temperature, wet bulb temperature, absolute humidity, mixing ratio, enthalpy |
Property | Description/Value |
---|---|
Cable connector | 4-pin M8 (IEC 60947-5-2) |
Materials | |
Body | PC/ABS blend |
Grid filter | PC (glass reinforced) |
Cable | Polyurethane or FEP |
Weight | |
Probe | 9 g (0.3 oz) |
Probe with 0.3 m (1 ft) cable | 20 g (0.7 oz) |
1 Dew point depression = ambient temperature − dew point
2 Not applicable with the plastic grid
filter.
3 Use lowest available operating voltage to minimize
heating.