With digital output, use the RHLIMIT command to set the maximum RH reading from 100 % (default) up to 120 %.
With analog output, the RHLIMIT command allows you to extend the high limit of the analog output scaling up to 120 %. This command does not change the scaling automatically. To change the scaling, use the ASEL command.
RHLIMIT [max_rh]<cr>
Syntax | Description |
---|---|
max_rh | Maximum reading of the RH parameter. Possible values are 100.0 … 120.0. The default value is 100.0. |
Example (extend the maximum RH reading to 120 %):
rhlimit 120
Max. RH % : 120.0 %RH
Example (on analog output, extend the maximum RH reading to 120 % and scale the RH output on channel 2 from (0 … 1 V) 0 … 100 % to (0 … 1 V) 0 … 120 %):
rhlimit 120
Max. RH % : 120.0 %RH
asel ?
Ch1 T lo : -20.00 'C ?
Ch1 T hi : 80.00 'C ?
Ch2 RH lo : 0.00 %RH ?
Ch2 RH hi : 100.00 %RH ? 120