Extend maximum RH reading - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide

With digital output, use the RHLIMIT command to set the maximum RH reading from 100 % (default) up to 120 %.

With analog output, the RHLIMIT command allows you to extend the high limit of the analog output scaling up to 120 %. This command does not change the scaling automatically. To change the scaling, use the ASEL command.

RHLIMIT [max_rh]<cr>

Syntax Description
max_rh Maximum reading of the RH parameter. Possible values are 100.0 … 120.0. The default value is 100.0.

Example (extend the maximum RH reading to 120 %):

rhlimit 120
Max. RH %      :    120.0 %RH

Example (on analog output, extend the maximum RH reading to 120 % and scale the RH output on channel 2 from (0 … 1 V) 0 … 100 % to (0 … 1 V) 0 … 120 %):

rhlimit 120
Max. RH %      :    120.0 %RH
asel ?
Ch1 T    lo    :   -20.00 'C ?
Ch1 T    hi    :    80.00 'C ? 
Ch2 RH   lo    :     0.00 %RH ?
Ch2 RH   hi    :   100.00 %RH ? 120