Status registers - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide
Table 1. Modbus status data registers (read-only)
Register number Address Register description Data format Note
513 0200hex Error status 16-bit integer

0000hex: One or more errors active

0001hex: No errors

516 0203hex Error code 32-bit integer
0204hex
518 0205hex Security hash 32-bit integer Security hash changes when any change is made to device settings or adjustments, but also returns back to the previous value if such changes are reverted completely.
0206hex
Table 2. Error codes on Modbus interface
Error code register value Corresponding error
1 Temperature measurement error. [44]
2 Humidity measurement error. [45]
4 Humidity sensor failure. [46]
8 Capacitance reference error. [47]
16 Ambient temperature out of range. [48]
32 Firmware checksum mismatch. [49]
64 Device settings corrupted. [50]
128 Additional configuration settings corrupted. [51]
256 Sensor coefficients corrupted. [52]
512 Main configuration settings corrupted. [53]
2048 Supply voltage out of range. [55]
8192 Non-volatile memory read/write failure. [57]
16384 Calibration certificate checksum mismatch. [58]

If several errors are active at the same time, the Modbus error code register contains the sum of currently active error codes.

Refer to Error messages in Insight software for more information about the error states.