HMP60 specifications - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide
Table 1. HMP60 measurement performance
Property Description/Value
Relative humidity
Measurement range 0 … 100 %RH
Typical accuracy:

at 0 … +40 °C (+32 … +140 °F)

±3 %RH (0 … 90 %RH)

±5 %RH (90 … 100 %RH)

at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F and +104 … +140 °F)

±5 %RH (0 … 90 %RH)

±7 %RH (90 … 100 %RH)

Humidity sensor Vaisala INTERCAP®
Temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Accuracy:

at +10 … +30 °C (+50 … +86 °F)

at −40 … +10 and +30 … +60 °C (−40 … +50 and +86 … +140 °F)

±0.5 °C (±0.9 °F)

±0.6 °C (±1.08 °F)

Dew point temperature
Measurement range −40 … +60 °C (−40 … +140 °F)
Typical accuracy:

at +32 … +40 °C (0 … +140 °F) when dew point depression 1 < +15 °C (< +59 °F)

±2 °C (±3.6 °F)

at −40 … 0 °C and +40 … +60 °C (−40 … +32 °F and +104 … +140 °F) when dew point depression < +10 °C (< +50 °F)

±3 °C (±5.4 °F)

Analog outputs
Accuracy at +20 °C (+68 °F) ±0.2 % of FS
Temperature dependence ±0.01 % of FS/°C (±0.006 % of FS/°F)
Table 2. HMP60 calculated parameters
Property Description/Value
Measurement ranges
Dew point temperature −40 … +60 °C (−40 … +140 °F)
Wet bulb temperature −40 … +60 °C (−40 … +140 °F)
Absolute humidity 0 … 130 g/m3 (0 … 57 gr/ft3)
Mixing ratio 0 … 153 g/kg (0 … 1069 gr/lb)
Enthalpy −40 … 459 kJ/kg (−9.6 … 207 BTU/lb)
Table 3. HMP60 operating environment
Property Description/Value
Operating temperature −40 … +60 °C (−40 … +140 °F)
EMC compliance EN 61326-1, industrial environment
Table 4. HMP60 inputs and outputs
Property Description/Value
Power consumption 1 mA average, max. peak 5 mA
Operating voltage 2
With 1 V / 2.5 V output 5 … 28 VDC
With 5 V output 8 … 28 VDC
With loop power converter 8 … 28 VDC
With digital output 5 … 28 VDC
Start-up time
Probes with analog output

4 s at operating voltage 13.5 … 16.5 VDC

2 s at other valid operating voltages

Probes with digital output 1 s
Outputs
2 channels 0 … 1 VDC / 0 … 2.5 VDC / 0 … 5 VDC / 1 … 5 VDC
1-channel loop-power converter (separate module, compatible with humidity accuracy only) 4 … 20 mA
Digital output (optional) RS-485 2-wire half duplex, supports Modbus RTU
External loads
0 … 1 V RL min. 10 kΩ
0 … 2.5 V /0 … 5 V RL min. 50 kΩ
Output parameters
Relative humidity, temperature, dew point temperature, wet bulb temperature, absolute humidity, mixing ratio, enthalpy
Table 5. HMP60 mechanical specifications
Property Description/Value
IP rating IP65 3
Body thread MI2x1 / 10 mm (0.4 in)
Cable connector 4-pin M8 (IEC 60947-5-2)
Materials
Body Stainless steel (AISI 316)
Grid filter Chrome coated ABS plastic
Cable Polyurethane or FEP
Weight
Probe 17 g (0.6 oz)
Probe with 0.3 m (1 ft) cable 28 g (1 oz)
1 Dew point depression = ambient temperature − dew point.
2 Use lowest available operating voltage to minimize heating.
3 Applicable with stainless steel sintered filter and PTFE sintered filter only.