Periodic maintenance - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide

The humidity measurement accuracy of the HMP60 and HMP110 series probes should be calibrated yearly. When calibration indicates that accuracy is not within specification:

  • HMP60 and HMP63: change the INTERCAP® sensor.
  • HMP110 and HMP113: adjust the measurement yourself, or have it adjusted at Vaisala.

Light cleaning of the probe and replacement of the filter should be done only when necessary.