To perform a 2-point adjustment to the capacitance measurement of the HMP110 or HMP113 using serial line you need:
- Vaisala USB cable (item code 219690)
- PC with a terminal application
- Two humidity references. The first point requires a < 50 %RH humidity reference, the second point must be > 50 %RH. There must also be
at least 30 %RH difference between the references.
The following example procedure uses the
HMK15 Humidity Calibrator. LiCl salt (11.3 %RH) at 25 °C is used as the first reference
point, NaCl (75.3 %RH at 25 °C) as the second.
-
Connect the USB cable to the PC, but do not connect it to the probe yet.
-
Remove the filter from the probe and insert the probe in the LiCl salt chamber of the
humidity calibrator (11.3 %RH).
-
Start a terminal application and set the correct connection settings. The default serial settings are 19200 8 N 1. Remember to check which COM port the USB cable is using.
-
Wait for 20 … 40 minutes for the humidity to stabilize.
-
Start the terminal session and connect the USB cable to the probe. If your probe is in the analog output mode, you need to press ENTER a few times to start it in
RS-485 mode.
-
Verify that the connection works by giving the ? command.
?
If the probe does not respond with device information:
- Disconnect the USB cable from the probe and retry. If your probe is in the analog output mode, press ENTER a few
times immediately after connecting the USB cable.
- Try the ?? command in case the probe is in POLL mode, open the line using the OPEN command if necessary.
- Check your serial line settings and cable connections.
-
When your serial connection is working, use the L command to see the current user adjustment parameters.
-
Use the ERRS command to see that no errors are active.
-
Use the SEND command to verify the currently measured RH value.
send
T= 22.9 'C RH= 11.1 %RH Td= -8.0 'C
-
Give the CRH command to display the current RH measurement. Press
ENTER to refresh the RH measurement to check that the measurement
has stabilized.
crh
RH: 11.39800112 Ref1 ?
RH: 11.38377201 Ref1 ?
RH: 11.37988329 Ref1 ?
-
When the shown RH measurement shown is stable, enter the value of the first measurement
humidity reference (11.3 %RH) and press ENTER.
crh
RH: 11.37988329 Ref1 11.3
Press any key when ready
>
-
After entering the first reference point correction, press ENTER
to continue to the second reference point correction.
-
Insert the probe in the NaCl salt chamber (75.3 %RH) and wait for 20 … 40 minutes for
humidity and temperature to stabilize.
-
Press ENTER to refresh the RH measurement to check that the
measurement has stabilized.
RH: 74.87782214 Ref2 ?
RH: 74.89157890 Ref2 ?
RH: 74.91556023 Ref2 ?
-
When the shown RH measurement shown is stable, enter the value of the second humidity
measurement reference (75.3 %RH) and press ENTER.
crh
RH: 11.37988329 Ref2 75.3
Press any key when ready
>
-
Check with the L command that the user adjustment parameters have
changed.
-
Use the SEND command to verify that the measurement is OK.
| If the entered adjustment is incorrect, you can remove the effects of
RH calibration (returning the RH measurement of the probe to the factory calibrated
state) with the CRHCLR command. For instructions, see
Clear adjustment of RH measurement. |