Display the currently active errors - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide

Use the ERRS command to display the currently active error codes.

ERRS<cr>

Example (no active errors):

errs
0000h
No errors
Table 1. Error messages on ERRS command
ERRS command response Corresponding error
T MEAS error Temperature measurement error. [44]
F meas error Humidity measurement error. [45]
RH sensor failure Humidity sensor failure. [46]
Frequency measurement outside the permissible value range Capacitance reference error. [47]
Ambient temperature error Ambient temperature out of range. [48]
Program flash check sum error Firmware checksum mismatch. [49]
Parameter flash check sum error Device settings corrupted. [50]
INFOA check sum error Additional configuration settings corrupted. [51]
SCOEFS check sum error Sensor coefficients corrupted. [52]
CURRENT check sum error Main configuration settings corrupted. [53]
Voltage error Supply voltage out of range. [55]
General flash failure w/r Non-volatile memory read/write failure. [57]
Calibration certificate check sum failure Calibration certificate checksum mismatch. [58]

Refer to Error messages in Insight software for more information about the error states.