HMP110 specifications - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide
Table 1. HMP110 measurement performance
Property Description/Value
Relative humidity
Measurement range 0 … 100 %RH
Accuracy: 1 2

at 0 … +40 °C (+32 … +104 °F)

±1.5 %RH (0 … 90 %RH)

±2.5 %RH (90 … 100 %RH)

at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F)

±3.0 %RH (0 … 90 %RH)

±4.0 %RH (90 … 100 %RH)

Factory calibration uncertainty at +20 °C (+68 °F)

±1.1 %RH (0 … 90 %RH)

±1.8 %RH (90 … 100 %RH)

Humidity sensor types

HUMICAP® 180R

HUMICAP® 180V

Stability ±2 %RH over 2 years
Temperature
Measurement range −40 … +80 °C (−40 … +176 °F)
Accuracy (probes with analog output):
at 0 … +40 °C (+32 … +104 °F) ±0.2 °C (±0.36 °F)
at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F) ±0.4 °C (±0.72 °F)
Accuracy (probes with digital output):
at +15 … +25 °C (+59 … +77 °F) ±0.1 °C (±0.18 °F)

at 0 … +15 °C (+ 32 … +59 °F) and +25 … +40 °C (+77 … +104 °F)

±0.15 °C (±0.27 °F)
at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F) ±0.4 °C (±0.72 °F)
Temperature sensor Pt1000 RTD Class F0.1 IEC 60751
Dew point temperature
Measurement range −40 … +80 °C (−40 … +176 °F)
Accuracy (at 20 °C (68 °F) and 80 %RH 1):

at 0 … +40 °C (+32 … +104 °F)

when dew point depression < +15 °C (+59 °F) 3

when dew point depression +15 … +25 °C (+59 … +77 °F)

 

±1 °C (±1.8 °F)

±2 °C (±3.6 °F)

at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F), when dew point depression < +15 °C (+59 °F)

±2 °C (±3.6 °F)
Analog outputs
Accuracy at +20 °C (+68 °F) ±0.2 % of FS
Temperature dependence ±0.01 % of FS/°C (±0.006 % of FS/°F)
Table 2. HMP110 calculated parameters
Property Description/Value
Measurement ranges
Dew point temperature −40 … +80 °C (−40 … +176 °F)
Wet bulb temperature −40 … +80°C (−40 … +176 °F)
Absolute humidity 0 … 291 g/m3 (0 … 127 gr/ft3)
Mixing ratio 0 … 548 g/kg (0 … 3836 gr/lb)
Enthalpy −40 … 1530 kJ/kg (−9.6 … 674 BTU/lb)
Table 3. HMP110 operating environment
Property Description/Value
Operating temperature −40 … +80 °C (−40 … +176 °F)
IP rating IP65 4

EMC compatibility

EN 61326‑1, industrial environment

Table 4. HMP110 inputs and outputs
Property Description/Value
Power consumption 1 mA average, max. peak 5 mA
Operating voltage 5
With 1 V / 2.5 V output 5 … 28 V DC
With 5 V output 8 … 28 V DC
With loop power converter 8 … 28 V DC
With digital output 5 … 28 V DC
Start‑up time
HMP110 probes with analog output

4 s at operating voltage 13.5 … 16.5 V DC

2 s at other valid operating voltages

HMP110 probes with digital output 1 s
Outputs
2 channels 0 … 1 V DC / 0 … 2.5 V DC / 0 … 5 V DC / 1 … 5 V DC
1‑channel loop-power converter (separate module, compatible with humidity accuracy only) 4 … 20 mA
Digital output (HMP110 probes with digital output) RS‑485 2‑wire half duplex, supports Modbus RTU
External loads
0 … 1 V RL min. 10 kΩ
0 … 2.5 V /0 … 5 V RL min. 50 kΩ
Output parameters
Relative humidity, temperature, dew point temperature, wet bulb temperature, absolute humidity, mixing ratio, enthalpy
Table 5. HMP110 mechanical specifications
Property Description/Value
Body thread MI2x1 / 10 mm (0.4 in)
Cable connector M8 4‑pin female (IEC 60947‑5‑2)
Materials
Body Stainless steel (AISI 316)
Grid filter Chrome coated ABS plastic
Cable Polyurethane or FEP
Weight
Probe 17 g (0.6 oz)
Probe with 0.3 m (1 ft) cable 28 g (1 oz)
1 Including non-linearity, hysteresis, and repeatability.
2 With HUMICAP® 180V sensor, accuracy is not specified below −20 °C (−4 °F) operating temperature.
3 Dew point depression = ambient temperature − dew point.
4 Applicable with stainless steel sintered filter and PTFE sintered filter only.
5 Use lowest available operating voltage to minimize heating.