Property | Description/Value |
---|---|
Relative humidity | |
Measurement range | 0 … 100 %RH |
Accuracy: 1 2 | |
at 0 … +40 °C (+32 … +104 °F) |
±1.5 %RH (0 … 90 %RH) ±2.5 %RH (90 … 100 %RH) |
at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F) |
±3.0 %RH (0 … 90 %RH) ±4.0 %RH (90 … 100 %RH) |
Factory calibration uncertainty at +20 °C (+68 °F) |
±1.1 %RH (0 … 90 %RH) ±1.8 %RH (90 … 100 %RH) |
Humidity sensor types |
HUMICAP® 180R HUMICAP® 180V |
Stability | ±2 %RH over 2 years |
Temperature | |
Measurement range | −40 … +80 °C (−40 … +176 °F) |
Accuracy (probes with analog output): | |
at 0 … +40 °C (+32 … +104 °F) | ±0.2 °C (±0.36 °F) |
at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F) | ±0.4 °C (±0.72 °F) |
Accuracy (probes with digital output): | |
at +15 … +25 °C (+59 … +77 °F) | ±0.1 °C (±0.18 °F) |
at 0 … +15 °C (+ 32 … +59 °F) and +25 … +40 °C (+77 … +104 °F) |
±0.15 °C (±0.27 °F) |
at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F) | ±0.4 °C (±0.72 °F) |
Temperature sensor | Pt1000 RTD Class F0.1 IEC 60751 |
Dew point temperature | |
Measurement range | −40 … +80 °C (−40 … +176 °F) |
Accuracy (at 20 °C (68 °F) and 80 %RH 1): | |
at 0 … +40 °C (+32 … +104 °F) when dew point depression < +15 °C (+59 °F) 3 when dew point depression +15 … +25 °C (+59 … +77 °F) |
±1 °C (±1.8 °F) ±2 °C (±3.6 °F) |
at −40 … 0 °C (−40 … +32 °F) and +40 … +80 °C (+104 … +176 °F), when dew point depression < +15 °C (+59 °F) |
±2 °C (±3.6 °F) |
Analog outputs | |
Accuracy at +20 °C (+68 °F) | ±0.2 % of FS |
Temperature dependence | ±0.01 % of FS/°C (±0.006 % of FS/°F) |
Property | Description/Value |
---|---|
Measurement ranges | |
Dew point temperature | −40 … +80 °C (−40 … +176 °F) |
Wet bulb temperature | −40 … +80°C (−40 … +176 °F) |
Absolute humidity | 0 … 291 g/m3 (0 … 127 gr/ft3) |
Mixing ratio | 0 … 548 g/kg (0 … 3836 gr/lb) |
Enthalpy | −40 … 1530 kJ/kg (−9.6 … 674 BTU/lb) |
Property | Description/Value |
---|---|
Operating temperature | −40 … +80 °C (−40 … +176 °F) |
IP rating | IP65 4 |
EMC compatibility |
EN 61326‑1, industrial environment |
Property | Description/Value |
---|---|
Power consumption | 1 mA average, max. peak 5 mA |
Operating voltage 5 | |
With 1 V / 2.5 V output | 5 … 28 V DC |
With 5 V output | 8 … 28 V DC |
With loop power converter | 8 … 28 V DC |
With digital output | 5 … 28 V DC |
Start‑up time | |
HMP110 probes with analog output |
4 s at operating voltage 13.5 … 16.5 V DC 2 s at other valid operating voltages |
HMP110 probes with digital output | 1 s |
Outputs | |
2 channels | 0 … 1 V DC / 0 … 2.5 V DC / 0 … 5 V DC / 1 … 5 V DC |
1‑channel loop-power converter (separate module, compatible with humidity accuracy only) | 4 … 20 mA |
Digital output (HMP110 probes with digital output) | RS‑485 2‑wire half duplex, supports Modbus RTU |
External loads | |
0 … 1 V | RL min. 10 kΩ |
0 … 2.5 V /0 … 5 V | RL min. 50 kΩ |
Output parameters | |
Relative humidity, temperature, dew point temperature, wet bulb temperature, absolute humidity, mixing ratio, enthalpy |
Property | Description/Value |
---|---|
Body thread | MI2x1 / 10 mm (0.4 in) |
Cable connector | M8 4‑pin female (IEC 60947‑5‑2) |
Materials | |
Body | Stainless steel (AISI 316) |
Grid filter | Chrome coated ABS plastic |
Cable | Polyurethane or FEP |
Weight | |
Probe | 17 g (0.6 oz) |
Probe with 0.3 m (1 ft) cable | 28 g (1 oz) |