Changing the INTERCAP sensor (HMP60 and HMP63) - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide

This procedure restores the humidity measurement accuracy of the probe. No adjustment after the sensor change is needed.

To perform this procedure, you need a new INTERCAP® sensor. It is also recommended that you replace the filter with a new one.

  1. Remove the filter from the probe by turning it counter-clockwise.
  2. There are 2 sensors under the filter, the INTERCAP® sensor and a temperature sensor. Identify the INTERCAP® sensor. Do not touch the temperature sensor.
  3. Pull out the old INTERCAP® sensor and insert a new one. Handle the new sensor by the plastic frame.
    CAUTION Do not touch the sensor plate.
  4. Attach a new filter on the probe.