Set analog output error indication level - HMP110 - HMP113 - HMP60 - HMP63

HMP60 and HMP110 Series User Guide

Document code
M211060EN
Revision
L
Language
English
Product
HMP110
HMP113
HMP60
HMP63
Document type
User guide

If the device is malfunctioning, the analog output is set to a specified level. This overrides the normal measurement output of the channel. The default error level is 0 V, or another value predefined by the customer when ordering the device. You can set the level using the AERR command. This command is not in use in the digital-only HMP110 probe option.

AERR [ch1 ch2] <cr>

Syntax Description
ch1 Error level of the analog output for channel 1. The available range depends on the output mode (check with AMODE command).
ch2 Error level of the analog output for channel 2. The available range depends on the output mode (check with AMODE command).

Example (show present output modes):

amode
Ch1 output     :  0 ... 1 V
Ch2 output     :  0 ... 1 V

Example (check present analog output error level):

aerr
Ch1 error out: 0.000V ?
Ch2 error out: 0.000V ?

Example (set analog output error level to 1 V on both channels):

aerr 1 1
Ch1 error out: 1.000V ?
Ch2 error out: 1.000V ?
The error output value is displayed only when there are minor electrical faults such as humidity sensor damage. When there is a severe device malfunction, the error output value is not necessarily shown.