You can issue all commands either in uppercase or lowercase. In the command examples, the keyboard input by the user is in bold type.
The notation <cr> refers to pressing the carriage return (ENTER key on your computer keyboard). Press ESC to clear the command buffer before starting to enter commands.
Command | Description |
---|---|
? | Output information about the device |
AERR | Set analog output error level |
AMODE | View or set the analog output mode |
AOVER [ON/OFF] | Allow analog outputs to exceed their range 10 % |
ASEL | Set analog output parameters and scaling |
CDATE | View or set the calibration date |
CODE | View the order code of the probe |
CRH | Calibrate and adjust RH measurement |
CRHCLR | Clear adjustment of RH measurement |
CT | Calibrate and adjust T measurement |
CTCLR | Clear adjustment of T measurement |
CTEXT | View or set the calibration information field |
ERRS | List present probe errors |
FILT [0.001 ... 1] | Set the result filtering |
FRESTORE | Restore factory settings |
HELP | List available commands |
INTV [0 ... 255 S/MIN/H] | Set the continuous output interval (for RUN mode) |
L | Displays user adjustment parameters |
R | Start the continuous outputting |
RESET | Reset the probe |
RHLIMIT | Extend maximum RH reading |
S | Stop the continuous outputting |
SDELAY [0 ... 255] | View or set serial line answer minimum delay |
SEND [0 ... 255] | Output the reading once |
SNUM | View the serial number of the probe |
UNIT | Select metric or non-metric output units |
VERS | View software version of the probe |
Command | Description |
---|---|
?? | Output information about the device in POLL mode |
ADDR [0 ... 255] | Set the probe address (for POLL mode) |
CLOSE | Close the temporary connection (Back to POLL mode) |
OPEN [0 ... 255] | Open a temporary connection to a POLL mode device |
SERI [baud p d s] | User Port settings (Default: 19200 N 8 1) baud: 300 ... 57600 |
SMODE [STOP/RUN/POLL/MODBUS/VDIGI/ANALOG] | Set the serial interface mode |