You can calibrate and adjust the probe of
your RFL100 data logger in 1 or 2 points using the MI70 indicator. You can also do the
1-point calibration so that you compare the reading of the probe to any MI70-compatible
Vaisala probe that provides the same measurement parameter.
-
Disconnect the probe from the data logger.
-
Connect the probe to be calibrated to port I of the MI70 measurement indicator using
the connection cable.
-
If you want to calibrate by comparing to the reading of a reference probe, connect it
to port II of the MI70 indicator.
-
Turn on the MI70 indicator.
-
Start the adjustment sequence from .
If you have 2 probes connected, make sure to start the adjustment sequence for probe
I.
-
MI70 notifies you that automatic power off is disabled during adjustment mode, select
OK to acknowledge.
-
Select RH or T parameter for adjustment and
select OK. This procedure assumes you are adjusting relative humidity, but the same principles
apply for temperature adjustment.
-
Insert the probe to be calibrated in the reference environment. If you are calibrating
relative humidity using 2 reference environments, use the dry reference first. If you have
a reference probe, insert that in the same environment.
If you are comparing against the reading of a reference probe, you can also use the
ambient condition as the reference environment, as long as its conditions are
stable.
-
Wait for the measurement to stabilize. You can follow the stabilization from the
GRAPH display. Select ADJUST when the reading is stabilized in the reference.
-
To perform the adjustment using 1 reference environment (1-point adjustment), perform
these steps:
-
Select .
-
When the measurement is stable, select READY.
-
Give the reference RH value by using the arrow buttons and select OK.
-
To confirm the adjustment, select YES. If you select NO, you return to the adjustment mode display and no changes are made.
-
Continue from 13.
-
To perform the adjustment using 2 reference environments (2-point adjustment), perform
these steps:
-
Select .
-
When the measurement is stable, select READY.
-
Give the reference RH value by using the arrow buttons and select OK.
-
Insert the probe to be calibrated in the second reference environment.
-
When the measurement is stable, select READY.
-
Give the reference RH value by using the arrow buttons and select OK.
-
To confirm the adjustment, select YES. If you select NO, you return to the adjustment mode display and no changes are made.
-
Continue from 13.
-
To perform the adjustment using a reference probe, perform these steps:
-
Select To same as RH[II].
-
To confirm the adjustment, select YES. If you select NO, you return to the adjustment mode display and no changes are made.
-
Calibration and adjustment is now completed. Select BACK to exit the adjustment mode and EXIT to return to the basic display.
-
Disconnect the calibrated probe from the MI70 indicator.
-
Reconnect the probe to the data logger.