Field checking using a reference instrument - RFL100
RFL100 User Guide
- Document code
- M211861EN
- Revision
- J
- Language
- English
- Product
- RFL100
- Document type
- User guide
You can perform a field check of the
RFL100 using any humidity and temperature measurement instrument with a display. Typically
field checking is done with a recently calibrated portable instrument.
-
Place the probe of the reference instrument (or the entire instrument) in the same
environment as the probe of the RFL100. The environment should be as stable as
possible.
-
Wait for 30 minutes for humidity and temperature to stabilize. Verify that measurements
are no longer changing at the end of the stabilization period.
-
Record the readings from both instruments.