Use the SERI command to set the serial line settings. The new settings will be taken into use when the transmitter is reset or powered up. Note that the transmitter will use its default serial line settings if it is in error state; see Error state.

Also note that changing the serial line settings will slightly alter how the transmitter works with the MI70 indicator. If you are not using the default settings of 19200 N 8 1, you will have to power off MI70 before it can find the transmitter (you cannot connect it on-the-fly).

Table 1. SERI command
Syntax Description
SERI [b p d s]<cr>
B
Baud rate (300, 600, 1200, 2400, 4800, 9600, or 19200)
P
Parity (n = none, e = even, o = odd)
D
Data bits (7 or 8)
S
Stop bits (1 or 2)

Example:

seri 9600 e 7 1