The STAT command displays the status of the transmitter in 4 characters.
The characters in the output have the following meaning:
[sensor purge] - [output frozen] - [settling] - [status]
During normal measurement all characters are zeroes. When purge is activated, output frozen, or the transmitter is in the settling period after sensor purge, the corresponding status character is 1.
The final status character can have several values:
- 0
- Normal status
- 1
- Transmitter error
- 7
- Retrying auto-calibration due to fluctuations in the supply voltage
- 9
- Transmitter startup
Some example status codes:
- 0109
- Transmitter performing startup procedure
- 1100
- Sensor purge in progress, output frozen
- 1110
- Settling period after purge, output frozen
- 0000
- Normal measurement
Syntax | Description |
---|---|
STAT<cr> | Display the status of the transmitter in 4 characters. |
Example:
|
It is normal for measurement output to freeze temporarily if the measurement environment is changing rapidly. |