Select analog output parameter and scaling - DMT152

DMT152 User Guide

Document code
M210865EN
Revision
K
Language
English
Product
DMT152
Document type
User guide

Use the ASEL command to select which measurement parameters the analog channels output. See Table 1 for the selectable parameters. After entering the command, you will be prompted to enter the scaling of the parameters for both channels. Press the Enter key to accept the suggested scaling values.

Table 1. ASEL command
Syntax Description
ASEL [xxx yyy]<cr>
xxx
Analog channel 1 output parameter
yyy
Analog channel 2 output parameter

Example:

asel tdf ppm
Ch1 Tdf  lo    :   -80.00 'C ?
Ch1 Tdf  hi    :    20.00 'C ?
Ch2 ppm  lo    :     0.00 ppm ?
Ch2 ppm  hi    : 1000000.00 ppm ?