The DSC74C with an outlet coil is designed for the most critical measurements at atmospheric pressure. In the DSC74C, the coil is connected to the sampling cell outlet to protect the sensor from ambient humidity disturbing the measurement. The coil can also be used as a cooling coil for sampling from high temperature processes.

DSC74C parts are:

  • Sampling cell, thread G3/8"
  • Connection part with a needle valve and an integrated leak screw
  • Reducing nipple (thread adapter), G3/8" - G1/2"
  • Reducing adapter (thread adapter), G3/8" - G1/4"
  • Diffusion coil (for measurements in atmospheric pressure)
Figure 1. Default assembly of DSC74C
1
Gas goes in. The coil can also be used here.
2
Gas comes out
3
Coil
4
Valve
Figure 2. Alternative assembly of DSC74C (for tight spaces)
1
Gas comes out
2
Coil
3
Thread, max. size 7 mm
4
Gas goes in
5
Valve

The thread size cannot exceed 7 mm. Use the provided adapter to avoid damage to the transmitter.