Specifications - DMT152

DMT152 User Guide

Document code
M210865EN
Revision
K
Language
English
Product
DMT152
Document type
User guide
Table 1. Measurement performance
Property Description/Value
Sensor

Vaisala DRYCAP® 180U

Thin-film capacitive polymer sensor

Recommended calibration interval 2 years
Dew point temperature 1
Measurement range −80 … −20 °C (−112 … −4 °F) Td
Accuracy
−80 … −40 °C (−112 … −40 °F) ±2 °C (3.6 °F) Td
−40 … −20 °C (−40 … −4 °F) ±3 °C (5.4 °F) Td
Non-calibrated range −100 … +20 °C (−148 … +68 °F) Td
Typical response time 63 % [90 %] at a gas temperature of +20 °C (+68 °F) and pressure of 1 bar:
−20 … −80 °C Td 0.5 min [7.5 min]
−80 … −20 °C Td 2 s [5 s]
Typical long-term stability Better than 2 °C (3.6 °F) / year
Analog output scalings
Option 1 −80 … +20 °C (−112 … +68 °F) Td
Option 2 −100 … 0 °C (−148 … +32 °F) Td
Option 3 Free scaling
Concentration by volume (ppm)
Measurement range (typical) 0 … 500 ppm
Accuracy at +20 °C (+68 °F), 1013 mbar ±(0.2 ppm + 20 % of reading)
Figure 1. Accuracy over temperature range
Table 2. Operating environment
Property Description/Value
Temperature −40 … +70 °C (−40 … +158 °F)
Relative humidity 0 … 100 %RH (up to +20 °C / +68 °F)
Pressure 0 … 50 bar (725 psia)
Measurement environment

For air, nitrogen, argon, helium, and oxygen 2

Not suitable for measurements in hydrogen or pure carbon dioxide

Sample flow rate No effect on measurement accuracy
EMC compliance

EN61326-1, Industrial environment

 

Table 3. Inputs and outputs
Property Description/Value
Two analog outputs (scalable) 4 … 20 mA, 0 … 20 mA (3-wire) 0 … 5 V, 0 … 10 V
Digital output RS-485 (2-wire)
Alarm-level indication by analog signal User selectable
Purge information 5 V, 10 V, 20 mA, or LED
Accuracy of analog outputs ±0.01 V / ±0.01 mA
ON/OFF output available instead of 2nd analog output
Purge inactive 0 V or 4 mA
Purge active 5 V, 10 V, or 20 mA
OR exceeded Td or ppm limit User selectable
Typical temperature dependence
Voltage output ±0.001 % of span / °C
Current output ±0.005 % of span / °C
Operating voltage
RS-485 output 11 … 28 VDC 3
Voltage output 15  … 28 VDC 3
Current output 21 … 28 VDC
Supply current
Normal measurement 20 mA + load current
During self-diagnostics Max. 220 mA pulsed
Supply voltage fluctuation Max. 0.3 V
External load
Voltage output Min. 10 kΩ
Current output Max. 500 Ω
Table 4. Mechanical specifications
Property Description/Value
Housing material (wetted parts) AISI316L
Stainless steel mesh filter Filter body AISI303, mesh AISI316L, grade 18 μm
Mechanical connections ISO G1/2", NPT 1/2", UNF 3/4"- 16", UNF 5/8"-18"
IP rating IP66
Storage temperature range −40 … +80 °C (−40 … +176 °F)
Weight with different thread options

ISO thread: 190 g (6.70 oz)

NPT thread: 210 g (7.41 oz)

UNF thread: 186 g (6.56 oz)

Table 5. Output cable specifications
Property Specification
Connector Straight, moulded M8 female (IEC 60947-5-2)
Color Black
Number of contacts 4
Conductors 0.25 mm2 (× 4) / 0.00039 inch2 (× 4)
Insulation PVC
Cable diameter 5.0 mm / 0.2 in
Protection class IP67 / NEMA 4
Wire color code EN50044
1 When the dew point is below 0 °C, the transmitter outputs frost point for Td.
2 Consult Vaisala if other chemicals are present. Consider safety regulations with flammable gases.
3 For extended temp. down to −40 °C (−40 °F) or pressure up to 50 bar (725 psia), the supply voltage is 21 … 28 VDC.