Show device information - DMT152

DMT152 User Guide

Document code
M210865EN
Revision
K
Language
English
Product
DMT152
Document type
User guide

The ? command outputs a listing of device information.

Table 1. ? command
Syntax Description
?<cr> Display device information.
??<cr> Display device information in POLL mode.

Example:

?
?
DMT152/1.10
Serial number  : D1930026
Batch number   : D1330070
Module number  : ????????
Sensor number  : C1420000
Cal. date      : 20080512
Cal. info      : Vaisala/HEL
Serial mode    : STOP
Baud P D S     : 19200 N 8 1
Output interval:      1S
Address        :      0
Pressure       : 1.0132 bar
Filter         : 1.000
Ch1 output     : 0 ...  5 V
Ch2 output     : U ALARM
Ch1 Tdf  lo    :   -80.00 'C
Ch1 Tdf  hi    :    20.00 'C
Ch2 Tdf  lo    :     0.00 'C
Ch2 Tdf  hi    :     0.00 'C

If the transmitter is in POLL mode, but a connection has not been opened using the OPEN command, issue the ?? command. For a description of the serial interface modes, see Set serial line operating mode.