You can temporarily disable the sensor purge function using the XPUR command. Note that disabling the sensor purge will affect the measurement performance of the transmitter.
Syntax | Description |
---|---|
XPUR [ON/OFF]<cr> | ON = Enable sensor purge OFF = Disable sensor purge |
Example:
|
After giving the XPUR OFF command, the transmitter will not perform a sensor purge until one of the following
events takes place:
- Transmitter is reset or powered up
- Sensor purge is enabled using the XPUR ON command