The transmitter has error codes that can be displayed on the serial line using the ERRS command. See section Display currently active errors.
Error code | Cause | ||
---|---|---|---|
Ambient temperature error | Ambient temperature out of operating range (probably too hot). | ||
Autocal error | Transmitter unable to complete auto-calibration after several attempts. Can be caused by continuously changing measurement environment, problem with supply voltage, or transmitter malfunction. | ||
Voltage error | Supplied operating voltage is too low or high for reliable operation, or there is a problem with voltage measurement. | ||
Voltage too low error | Supplied operating voltage is too low for reliable operation. | ||
Voltage too low for mA output error |
Supplied operating voltage is too low for reliable operation of the analog current output.
|
||
T meas error | Internal error, can be caused by damage to the sensor. | ||
F meas error | Internal error, can be caused by damage to the sensor. | ||
Internal error | Transmitter unable to function. Can be caused by transmitter malfunction or unsuitable measurement environment. | ||
Parameter flash check sum error | Internal error | ||
Program flash check sum error | Internal error | ||
INFOA check sum error | Internal error | ||
SCOEFS check sum error | Internal error | ||
CURRENT check sum error | Internal error | ||
Unknown error | Internal error |