Calibrate analog output - DMT152

DMT152 User Guide

Document code
M210865EN
Revision
K
Language
English
Product
DMT152
Document type
User guide
Use the ACAL command to force the analog outputs to the following values:
  • Current output: 2 mA and 18 mA
  • Voltage output: 10 % and 90 % of the range

After entering the ACAL command, use a calibrated current/voltage meter to measure the output, and enter the values. Note that the calibration is only valid for the currently selected analog output modes.

Table 1. ACAL command
Syntax Description
ACAL [channel]<cr> [channel] = Analog output channel: 1 or 2

Example:

acal 1
Ch1: I1 ( mA  ) ? 0.166
     I2 ( mA  ) ? 18.1
      3.61058569E+02     1.40927863E+03