This procedure applies only to humidity probes HMP75B, HMP76B, and HMP77B. |
-
Open the menu by pressing
Open.
-
Select Functions and press
.
-
Select Chemical Purge and press
Start.
-
Confirm the purge by pressing
Yes. Chemical purge takes up to 6 minutes. During that time there are no readings on the display. A message appears every 15 seconds showing the time required to complete the chemical purge. The chemical purge icon
is shown on the upper left corner.
- When the chemical purge is complete the basic display with readings returns.
-
It may take a few minutes for the temperature reading to stabilize.
You can create a shortcut to the chemical purge function by setting a shortcut key to point to it. See Programming shortcut keys.