Carry out the steps listed in Making temperature adjustments before continuing with this adjustment option. |
The difference between the 2 references must be at least 30 °C or the adjustment cannot be done. |
-
Choose 2-point adjustment and press
Select.
-
Set the probe to the lower reference temperature. You can follow the stabilization from the graphical display by pressing
Graph. Press
Ready when the reading has stabilized.
-
Enter the correct lower reference temperature value with the arrow buttons and press
OK.
-
Set the probe to the higher reference temperature. You can follow the stabilization from the graphical display by pressing
Graph. Press
Ready when the reading has stabilized.
-
Enter the correct higher reference temperature value with the arrow buttons and press
OK.
-
Confirm the adjustment by pressing
Yes. If you press
No, you return to the adjustment mode display and no changes are made.
-
The adjustment is done. Press
Back and
Exit to return to the basic display.