Usually it is recommended to make an adjustment in 2 reference humidities. If you make an adjustment using 1 reference humidity (1-point adjustment), select the reference humidity so that it represents the measurement environment well.

Make the adjustment in 1 point as instructed in 2-point RH adjustment of HMP70 series probes (select 1-point adjustment instead of 2-point adjustment and follow the instructions on the display).