When measuring in pressurized environments, the actual process pressure value must be given to the HM70. The pressure can be given in the
following units:
- hPa: Absolute pressure given in the unit of hPa.
- barg: Gauge pressure given in the unit of bar. Indicates the pressure difference between normal atmospheric pressure and the actual process pressure.
- bara: Absolute process pressure given in the unit of bar.
- psia: Absolute pressure given in the unit of psi.
-
Open the menu by pressing
Open.
-
Select Environment and press
.
-
Press
Unit to change the pressure unit. Default unit is hPa.
-
Press
Set to set the ambient pressure value.
-
Set the pressure value with the arrow buttons. Press
+/- to change the sign of the pressure value. Press
OK to save the value.
-
To return to the basic display, press
Exit.