Register number | Address | Description | Data format | Unit |
---|---|---|---|---|
1 | 0000hex | Vaporized hydrogen peroxide concentration by volume | 32-bit float | ppm |
3 | 0002hex | H2O+H2O2 relative saturation | 32-bit float | %RS |
5 | 0004hex |
Temperature used for calculation. By default, shows the temperature from the temperature probe (register 0016hex). If temperature compensation is turned ON (register 0505hex), this register shows the volatile value for given temperature reading (register 0302hex). See Table 1. |
32-bit float | °C |
7 | 0006hex | Relative humidity | 32-bit float | %RH |
9 | 0008hex | Absolute hydrogen peroxide | 32-bit float | mg/m3 |
11 | 000Ahex | H2O+H2O2 dew point temperature | 32-bit float | °C |
15 | 000Ehex | Water concentration by volume | 32-bit float | ppm |
17 | 0010hex | Water vapor pressure | 32-bit float | hPa |
19 | 0012hex | Absolute humidity (water) | 32-bit float | g/m3 |
21 | 0014hex | Water vapor saturation pressure | 32-bit float | hPa |
23 | 0016hex | Temperature from the temperature probe | 32-bit float | °C |
25 | 0018hex | H2O2 vapor pressure | 32-bit float | hPa |
27 | 001Ahex | Water vapor saturation pressure (H2O+H2O2) | 32-bit float | hPa |
257 | 0100hex | Vaporized hydrogen peroxide concentration by volume | 16-bit signed integer | ppm |
258 | 0101hex | H2O+H2O2 relative saturation | 16-bit signed integer | %RS * 100 |
259 | 0102hex |
Temperature used for calculation. By default, shows the temperature from the temperature probe (register 010Bhex). If temperature compensation is turned ON (register 0505hex), this register shows the volatile value for given temperature reading (register 0302hex). See Table 1. |
16-bit signed integer | °C * 100 |
260 | 0103hex | Relative humidity | 16-bit signed integer | %RH *100 |
261 | 0104hex | Absolute hydrogen peroxide | 16-bit signed integer | mg/m3 |
262 | 0105hex | H2O+H2O2 dew point temperature | 16-bit signed integer | °C * 100 |
264 | 0107hex | Water concentration by volume | 16-bit signed integer | ppm |
265 | 0108hex | Water vapor pressure | 16-bit signed integer | hPa |
266 | 0109hex | Absolute humidity (water) | 16-bit signed integer | g/m3 |
267 | 010Ahex | Water vapor saturation pressure (H2O) | 16-bit signed integer | hPa |
268 | 010Bhex | Temperature from the temperature probe | 16-bit signed integer | °C * 100 |
269 | 010Chex | H2O2 vapor pressure | 16-bit signed integer | hPa |
270 | 010Dhex | Water vapor saturation pressure (H2O+H2O2) | 16-bit signed integer | hPa |