Measurement data registers - HPP272

HPP272 User Guide

Document code
M211972EN
Revision
G
Language
English
Product
HPP272
Document type
User guide
Table 1. Modbus measurement data registers (read-only)
Register number Address Description Data format Unit
1 0000hex Vaporized hydrogen peroxide concentration by volume 32-bit float ppm
3 0002hex H2O+H2O2 relative saturation 32-bit float %RS
5 0004hex

Temperature used for calculation. By default, shows the temperature from the temperature probe (register 0016hex).

If temperature compensation is turned ON (register 0505hex), this register shows the volatile value for given temperature reading (register 0302hex). See Table 1.

32-bit float °C
7 0006hex Relative humidity 32-bit float %RH
9 0008hex Absolute hydrogen peroxide 32-bit float mg/m3
11 000Ahex H2O+H2O2 dew point temperature 32-bit float °C
15 000Ehex Water concentration by volume 32-bit float ppm
17 0010hex Water vapor pressure 32-bit float hPa
19 0012hex Absolute humidity (water) 32-bit float g/m3
21 0014hex Water vapor saturation pressure 32-bit float hPa
23 0016hex Temperature from the temperature probe 32-bit float °C
25 0018hex H2O2 vapor pressure 32-bit float hPa
27 001Ahex Water vapor saturation pressure (H2O+H2O2) 32-bit float hPa
257 0100hex Vaporized hydrogen peroxide concentration by volume 16-bit signed integer ppm
258 0101hex H2O+H2O2 relative saturation 16-bit signed integer %RS * 100
259 0102hex

Temperature used for calculation. By default, shows the temperature from the temperature probe (register 010Bhex).

If temperature compensation is turned ON (register 0505hex), this register shows the volatile value for given temperature reading (register 0302hex). See Table 1.

16-bit signed integer °C * 100
260 0103hex Relative humidity 16-bit signed integer %RH *100
261 0104hex Absolute hydrogen peroxide 16-bit signed integer mg/m3
262 0105hex H2O+H2O2 dew point temperature 16-bit signed integer °C * 100
264 0107hex Water concentration by volume 16-bit signed integer ppm
265 0108hex Water vapor pressure 16-bit signed integer hPa
266 0109hex Absolute humidity (water) 16-bit signed integer g/m3
267 010Ahex Water vapor saturation pressure (H2O) 16-bit signed integer hPa
268 010Bhex Temperature from the temperature probe 16-bit signed integer °C * 100
269 010Chex H2O2 vapor pressure 16-bit signed integer hPa
270 010Dhex Water vapor saturation pressure (H2O+H2O2) 16-bit signed integer hPa