Performing manual temperature adjustment with Indigo80 handheld indicator - HPP272 - SWINSIGHT10

HPP272 User Guide

Document code
M211972EN
Revision
G
Language
English
Product
HPP272
Document type
User guide
  • Probe connection cable (cable 272075SP or flat cable CBL210493SP: see accessory information)
  • Reference environment(s) for producing the desired reference values

This procedure tells you how to use the indicator to adjust the probe's measurement in 1 ... 2 points.

  1. Make sure you have completed the RH for H2O2 calibration and adjustment.
  2. Prepare the temperature calibrator equipment for temperature calibration point 1.
  3. Connect the probe that you are adjusting to Indigo80.
  4. Perform a sensor purge, or wait for the start-up purge to finish.
  5. Open the main menu by pressing .
  6. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
  7. Select Calibrate / Adjust.
  8. Select Temperature.
  9. Place the probe to the first reference point and select Reference value, point 1 to open the stabilizing graph view. Wait for the measurement to stabilize fully.

    The stabilization typically takes at least 10 ... 15 minutes.

    1. When the reading has stabilized, press .
    2. Use the arrow buttons to set the value of the reference, and then select OK or Set.
    If you already have the reference and measured values available for your calibration point(s), you can just skip the stabilizing graph view with the button to get to the view where you can set the reference value. Then replace the automatically filled Measured value, point 1 and Measured temperature, point 1 with your actual values.
  10. If you want to adjust the measurement in 2 points, place the probe to the second reference point. Then scroll down the view and select Reference value, point 2 to open the stabilizing graph view. Wait for the measurement to stabilize fully.
    1. When the reading has stabilized, press .
    2. Use the arrow buttons to set the value of the second reference, and then select OK or Set.
  11. Check the difference between each reference and measured value. Very large differences may be due to insufficient stabilization time or unsuitable calibration setup.
    • To store the adjustments to the probe, scroll down and select Activate adjustment > OK. Confirm by selecting OK.
    • To exit without taking the adjustment in use, press to end calibration.
  12. Update the calibration information in Calibrate / Adjust > Calibration information. Navigate to the Calibration information menu by pressing .
  13. Press to end calibration.