Restoring factory adjustment with Indigo80 handheld indicator - HPP272

HPP272 User Guide

Document code
M211972EN
Revision
G
Language
English
Product
HPP272
Document type
User guide
  • Probe connection cable (cable 272075SP or flat cable CBL210493SP: see accessory information)

If needed, you can restore the factory adjustment for your probe using the Indigo80 handheld indicator.

  1. Connect the probe(s) to the indicator.
  2. Open the main menu by pressing .
  3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
  4. Select Calibrate / Adjust.
  5. Select any of the available parameters to access the view where you can restore the factory adjustment.
  6. Select Manual adjustment.
  7. Scroll down and select Restore factory adjustment > OK.
  8. Confirm by selecting OK.
  9. Press to end calibration.