Configuring environment settings with Indigo80 handheld indicator - HPP272

HPP272 User Guide

Document code
M211972EN
Revision
G
Language
English
Product
HPP272
Document type
User guide
  • Probe connection cable (cable 272075SP or flat cable CBL210493SP: see accessory information)

For optimal measurement accuracy, you can use the Indigo80 handheld indicator to give additional information about the connected probe’s measurement environment.

The environment settings available via Indigo80 correspond to the environmental compensation settings of the probe. Therefore, Environment settings and Settings > Compensation setpoints are interconnected: the configuration set in either menu is applied to both.
  1. Connect the probe(s) to the indicator.
  2. Open the main menu by pressing .
  3. Select Devices. If you have more than one device connected to the indicator, make a further selection between the devices.
  4. Select Environment settings to access the settings for your probe.

    The available environment settings depend on the probe and probe firmware version.

    Pressure

    Temporary value for pressure compensation. This value is used instead of the default pressure compensation value (1013.25 hPa) only if Pressure compensation on/off is ON. At device reset, compensation value changes back to the power-up default value. You can find the power-up defaults in Devices > Settings .

    Temperature

    Temporary value of the temperature that is used in the calculation of other measurement parameters. This value is used instead of the temperature probe measurement if Temperature compensation mode is Setpoint.

    At device reset, compensation value changes back to the power-up default value. You can find the power-up defaults in Devices Settings.

  5. Select a setting from the list, for example, Pressure.
    1. Select the unit, if prompted.
    2. Use the arrow buttons to set the desired value, and then select OK or Set.

      You can also select to read the value from another probe connected to the indicator, provided that the probe can output the required compensation parameter. Press to highlight the Read value from field, and then press to select the other probe as the source of the value.

  6. Scroll down the Environment settings view and check that the following settings apply:
    • For Pressure, Pressure compensation on/off is set to On.
    • For Temperature, Temperature compensation mode is Setpoint.
  7. Exit the menu by pressing or return to the main menu by pressing .