Configuring Ceilometer - BLVIEW - CL31 - CL51

BL‑View User Guide

Document code
M211185EN
Revision
E
Language
English
Product
BLVIEW
CL31
CL51
Document type
User guide

If you have not ordered the ceilometer together with BL-View, you may need to change the ceilometer settings before using BL-View to analyze live data from the ceilometer.

For optimum performance, use the recommended settings for the selected ceilometer type.
Setting Supported Values Recommended for CL31 Recommended for CL51
Data message

msg1_10x770

  • msg2_10x770
  • If the Sky Condition algorithm is disabled:

    msg1_10x770

  • msg2_10x1540
  • If the Sky Condition algorithm is disabled:

    msg1_10x1540

msg1_20x385

msg1_5x1500

msg1_10x1540

msg2_10x770

msg2_20x385

msg2_5x1500

msg2_10x1540

Message interval 2 … 75 seconds 16 seconds 36 seconds
  1. Open the connection to the ceilometer.
  2. Open the communication port to the ceilometer.
  3. To set advanced level commands, type:
    advanced
  4. To set the message format, type:
    set message type <message format>
    • For example, for CL31
      set message type msg2_10x770
    • For example, for CL51:
      set message type msg2_10x1540
  5. To set the message interval, type:
    • For CL31:
      set message interval 16
    • For CL51:
      set message interval 36
  6. To enable the range gate normalization, type:
    set message profile noise_h2 on

    The range gate data is always range-normalized, including the noise.

  7. To get the factory parameters of CL31, type:
    get params factory
  8. Check the receiver type for CL31 in the response. If the receiver type is CLR321, go to the next step. Otherwise, change the algorithm parameters with the following command:
    set factory algo_sensit high_cloud 5
  9. Close the communication port.
  10. Verify that the output message is correct.