Battery failure |
Battery is dead. |
Replace the battery. |
Battery voltage low |
The unit has been powered too long using the backup battery. |
Connect the line voltage or replace the battery. |
Battery fails to recharge. |
If the battery is old, replace it. If the battery is OK, check the
operation of ceilometer engine board CLE321. |
Blower failure |
Blower cable is not connected. |
Check that the blower cable is connected. |
Window blower circuit breaker is not ON. |
Check that the window blower circuit breaker is ON. |
Line voltage level is not correct. |
Check that the line voltage is present and correct. |
Blower is stuck. |
Check if visible obstacles disable blower operation. |
Blower is damaged. |
Replace the blower. |
Blower is blowing, but only cool air. The blower heater is
malfunctioning. |
Replace the blower. |
Ceilometer engine board failure |
Non-critical fault in ceilometer engine board CLE321. |
Replace ceilometer engine board CLE321. |
Heater fault |
Window blower circuit breaker is not ON. |
Check that the window blower circuit breaker is ON. |
Line voltage level is not correct. |
Check that the line voltage is present and correct. |
Heater is damaged. |
Replace the heater. |
High background radiance |
Direct sunlight. |
In case there are no alarms, the measurements are OK. If there is an
alarm, the measurements are invalid. |
High humidity (option) |
Relative humidity > 85 %. Water has been condensed inside the
ceilometer. Leakage in the enclosure or leakage in the door. |
Take the ceilometer inside, open the maintenance door and let it dry
in a warm air conditioned place. Condensed water on optical surface may disturb the
measurement. There is a danger of short circuits. |
Humidity sensor failure (option) |
Humidity sensor is not connected. |
Connect the humidity sensor or disable the option in the software. |
Humidity sensor is damaged. |
Contact technical support and send the ceilometer to Service Center
for the humidity sensor to be replaced. |
Laser monitoring failure |
Laser power monitor board CLM311 has failed. |
Replace CLM311. |
Laser power measurement on ceilometer engine board CLE321 has
failed. |
Replace CLE321. |
Receiver warning |
A non-critical CLR321 receiver failure. |
Replace CLR321. |
Tilt angle > 45 degrees warning |
The unit is not correctly installed or the tilt angle is larger than 45°. |
Check the installation. The height measurement accuracy is lost with tilt angles greater than 45°. |
Transmitter expires |
Laser diode is too old. |
Replace transmitter CLT321. The measurement data is valid, but it is
possible that some clouds are missed. |
Window contamination (appears from time to time and
lasts < 5 min) |
Usually drizzle drops on the window. |
The blower cleans the window. If clouds are detected, the height
information is correct. If no clouds are detected, it is possible that high clouds are
missed. |
Window contamination (continuous) |
For example, bird droppings, leaves, or dust have contaminated the
window. |
Clean the window. |
Working from battery |
The unit has battery power left for a maximum of
1 ... 2 hours. |
Connect the line voltage or replace the battery. |