You can modify the algorithm settings for
Merged
Gradient and profile fit, Gradient, and Profile data.
The modifications apply to that preset only. Other presets that use the same algorithm
remain unchanged.
| Saving a new or an edited preset closes
all open plots except the live default plot and resets the view. |
-
In BL-View, select .
-
Select a preset from the list on the left.
| You cannot modify the Default preset. |
-
In Algorithm method, select the algorithm to modify, and select Edit.
-
In Advanced algorithm settings, modify the settings.
| The available settings depend on the
selected algorithm. |
-
For Gradient settings, set the Algorithm sensitivity.
The available range is 0 … 20. The default value is 5.
-
For Profile Fit settings, set the following:
- SNR
-
Set the signal-to-noise ratio.
The available range is 0 … 100. The default value is 1.
- SNR average window
-
Set the average window for the
signal-to-noise ratio.
The available range is 20 … 1000 m (66 ... 3280 ft). The default value is 500 m
(1640 ft).
- Night profile length
-
Set the length of the night
profile.
The available range is 10 … 4500 m
(33 ... 14 760 ft). The default value is 1000 m (3280 ft).
- Day profile length
-
Set the length of the day
profile.
The available range is 10 … 4500 m
(33 ... 14 760 ft). The default value is 4000 m (13 120 ft).
- Outlier removal strength
-
Set the outlier removal
strength.
The available range is 0 … 500. The default value is 60.
-
Select OK.
-
Select Save and Close.
The changes are reflected in the Information panel for those plots that use the modified preset.
To use the modified preset, select