Configuring Message Settings - BLVIEW - CL31 - CL51

BL‑View User Guide

Document code
M211185EN
Revision
E
Language
English
Product
BLVIEW
CL31
CL51
Document type
User guide
  1. In BL-View, select System > Settings > Devices.
  2. In Messages, set the Message type.

    This value must match the message format set for the ceilometer.

    Setting Supported Values Recommended for CL31 Recommended for CL51
    Data message

    msg1_10x770

    • msg2_10x770
    • If the Sky Condition algorithm is disabled:

      msg1_10x770

    • msg2_10x1540
    • If the Sky Condition algorithm is disabled:

      msg1_10x1540

    msg1_20x385

    msg1_5x1500

    msg1_10x1540

    msg2_10x770

    msg2_20x385

    msg2_5x1500

    msg2_10x1540

    Message interval 2 … 75 seconds 16 seconds 36 seconds
  3. In Message interval, set the message interval in seconds.
  4. Select Save and Close.