Modifying Calculation Presets - BLVIEW - CL31 - CL51

BL‑View User Guide

Document code
M211185EN
Revision
E
Language
English
Product
BLVIEW
CL31
CL51
Document type
User guide
You can modify values of calculation presets, reset presets to default values, and delete presets.
Saving a new or an edited preset closes all open plots except the live default plot and resets the view.
  1. In BL-View, select System > Settings > Presets.
  2. Select a preset from the list on the left.
    You cannot modify the Default preset.
  3. Select one of the following:
    • To reset the settings to default values, select Reset.
    • To delete a preset, select Delete.
    • To modify the algorithm settings, see Modifying Algorithm Settings of Calculation Presets.
    • To modify an existing preset, edit the values:
      Boundary layer min

      Set the lowest reported boundary layer height.

      The available range is 30 … 500 m (99 … 1640 ft). The default value is 30 m (99 ft).

      Boundary layer max

      Set the highest reported boundary layer height.

      The available range is 200 … 4000 m (656 ... 13 120 ft). The default value is 4000 m (13 120 ft).

      Number of boundary layers

      Set the maximum number of boundary layers shown in BL-View.

      The available range is 1 … 3. The default value is 3.

      Time averaging

      Select defined by User and set the time averaging interval.

      The default value is 30 minutes.

      Height averaging

      Select defined by User and set the height averaging interval.

      The default value is 360 m (1181 ft).

  4. Select Save and Close.

The changes are reflected in the Information panel for those plots that use the modified preset.