Use the serial line to test the operation of all the analog outputs. Use the ATEST command to force the analog outputs to entered values. The set values remain valid until you enter the ATEST command without parameters or reset the probe. Typing the ATEST command without parameters outputs the current output value.
ATEST V V
Parameter | Description |
---|---|
V | Output value for channel 1 (V) |
V | Output value for channel 2 (V) |
Example
> atest
2.412 6301
1.943 5090
>
Example
> atest 1.5 1.5
1.500 3948
1.500 3948
>