When adjusting the analog output the analog output is forced to the following values:
Voltage output: 10 % and 90 % of the range
To measure voltage, connect HMP155 to a calibrated voltage meter.
To change HMP155 to the adjustment mode, hold down the ADJ button until a green LED lights up. |
You cannot use the serial line command ACAL with HMP155 passive output version. |
Type the ACAL command and type the multimeter reading for each case. Continue by pressing ENTER. When a channel is specified, only the specified analog output channel is adjusted.