- Carry out chemical purge, if available.
- Uncover the adjustment buttons on the probe by removing the protective cover and the calibration seal.
-
Open the protective plug.
There are 3 buttons marked -, □, and +. There is also a 2-colored indicator LED. See Figure 1.
-
To change HMP155 to the adjustment mode, press the ADJ button
and hold it down until the green indicator LED lights up.
Do not touch the adjustment buttons before the conditions have stabilized. -
To perform the low humidity offset adjustment, remove the filter and insert the probe
into a measurement hole of the dry end reference chamber (for example, LiCl:
11 %RH).
Do not touch the adjustment buttons before the conditions have stabilized. This takes approximately 30 minutes. - Using the - and + buttons, make sure the Aout voltage is correct and press the ADJ button. The green indicator LED turns off and back on.
-
Insert the probe into the high-end reference chamber (for example, NaCl: 75 %RH chamber
in the humidity calibrator HMK15) and perform the high humidity gain adjustment by using
the - and + buttons to make sure the
Aout voltage is correct. To finish the relative humidity calibration and
adjustment, press the ADJ button and the red indicator LED lights
up.
If you do not want to perform the temperature adjustment at this time, press the ADJ button twice. The indicator LED turns off. After a reset, HMP155 exits the adjustment mode. Otherwise, continue with the next step. -
Insert the probe into a known reference temperature (if HMK15 humidity calibrator is
not used) and let the temperature reading stabilize.
Do not touch the adjustment buttons before the conditions have stabilized. -
Using the - and + buttons, make the
temperature offset adjustment by making sure the Aout voltage is correct and
press the ADJ button. Red indicator LED turns off and back.
If you do not wish to perform the two-point temperature adjustment now, press the ADJ button one more time so that the red indicator LED turns off. After reset, HMP155 exits the adjustment mode. Otherwise, continue with the next step. -
Insert the probe into another reference temperature.
Do not touch the adjustment buttons before the conditions have stabilized. - Using the - and + buttons, make the temperature gain adjustment by making sure the Aout voltage is correct.
- Press the ADJ button one more time so that the red LED turns off to indicate the probe has quit the adjustment mode.
-
After the procedure, tighten the protective cover properly.
To avoid water leaking in, make sure that there is no gap.