Use the FCRH command for relative humidity calibration after a sensor change.
>fcrh
RH : 25.19 Ref1 ? 11.3
Press any key when ready ...
RH : 70.02 Ref2 ? 75.5
OK
>
OK indicates that the adjustment has been successful.
Use the FCRH command for relative humidity calibration after a sensor change.
>fcrh
RH : 25.19 Ref1 ? 11.3
Press any key when ready ...
RH : 70.02 Ref2 ? 75.5
OK
>
OK indicates that the adjustment has been successful.